Ultrasound guided spine needle insertion. Chen, Elvis C. S.; Mousavi, Parvin; Gill, Sean; Fichtinger, Gabor; Abolmaesumi, Purang
Abstract
An ultrasound (US) guided, CT augmented, spine needle insertion navigational system is introduced. The system consists of an electromagnetic (EM) sensor, an US machine, and a preoperative CT volume of the patient anatomy. Three-dimensional (3D) US volume is reconstructed intraoperatively from a set of two-dimensional (2D) freehand US slices, and is coregistered with the preoperative CT. This allows the preoperative CT volume to be used in the intraoperative clinical coordinate. The spatial relationship between the patient anatomy, surgical tools, and the US transducer are tracked using the EM sensor, and are displayed with respect to the CT volume. The pose of the US transducer is used to interpolate the CT volume, providing the physician with a 2D "x-ray vision" to guide the needle insertion. Many of the system software components are GPU-accelerated, allowing real-time performance of the guidance system in a clinical setting. Copyright 2010 Society of Photo-Optical Instrumentation Engineers. One print or electronic copy may be made for personal use only. Systematic reproduction and distribution, duplication of any material in this paper for a fee or for commercial purposes, or modification of the content of the paper are prohibited.
Item Metadata
Title |
Ultrasound guided spine needle insertion.
|
Creator | |
Publisher |
Society of Photo-Optical Instrumentation Engineers (SPIE)
|
Date Issued |
2010
|
Description |
An ultrasound (US) guided, CT augmented, spine needle insertion navigational system is introduced. The system consists of an electromagnetic (EM) sensor,
an US machine, and a preoperative CT volume of the patient anatomy. Three-dimensional (3D) US volume is reconstructed intraoperatively from a set of
two-dimensional (2D) freehand US slices, and is coregistered with the preoperative CT. This allows the preoperative CT volume to be used in the intraoperative
clinical coordinate. The spatial relationship between the patient anatomy, surgical tools, and the US transducer are tracked using the EM sensor, and are
displayed with respect to the CT volume. The pose of the US transducer is used to interpolate the CT volume, providing the physician with a 2D "x-ray vision"
to guide the needle insertion. Many of the system software components are GPU-accelerated, allowing real-time performance of the guidance system in a clinical setting.
Copyright 2010 Society of Photo-Optical Instrumentation Engineers.
One print or electronic copy may be made for personal use only. Systematic reproduction and distribution,
duplication of any material in this paper for a fee or for commercial purposes, or modification of the content of the paper are prohibited.
|
Genre | |
Type | |
Language |
eng
|
Date Available |
2011-08-16
|
Provider |
Vancouver : University of British Columbia Library
|
Rights |
Attribution-NonCommercial-NoDerivatives 4.0 International
|
DOI |
10.14288/1.0107556
|
URI | |
Affiliation | |
Citation |
Chen, Elvis C. S.; Mousavi, Parvin; Gill, Sean; Fichtinger, Gabor; Abolmaesumi, Purang. Ultrasound guided spine needle insertion. Medical Imaging 2010: Visualization, Image-Guided Procedures, and Modeling, edited by Kenneth H. Wong, Michael I. Miga, Proceedings of SPIE, Volume 7625, 762538, 2010.
|
Publisher DOI |
10.1117/12.843716
|
Peer Review Status |
Reviewed
|
Scholarly Level |
Faculty
|
Copyright Holder |
Abolmaesumi, Purang
|
Rights URI | |
Aggregated Source Repository |
DSpace
|
Item Media
Item Citations and Data
Rights
Attribution-NonCommercial-NoDerivatives 4.0 International